Here is a quote from http://sunnyday.mit.edu/papers/therac.pdf:
"In addition, the Therac-25 software has more responsibility for maintaining safety than the software in the previous machines. The Therac-20 has independent protective circuits for monitoring the electron-beam scanning plus mechanical interlocks for policing the machine and ensuring safe operation. The Therac-25 relies more on software for these functions. AECL took advantage of the computer's abilities to control and monitor the hardware and decided not to duplicate all the existing hardware safety mechanisms and interlocks."
So, regarding these important safety aspects, even the Therac-20 was better than the Therac-25!
The linked post also mentions this:
"Preceding models used separate circuits to monitor radiation intensity, and hardware interlocks to ensure that spreading magnets were correctly positioned."
And indeed, the Therac-20 also had the same software error as the Therac-25! However, quoting again from the paper:
"The software error is just a nuisance on the Therac-20 because this machine has independent hardware protective circuits for monitoring the electron beam scanning. The protective circuits do not allow the beam to turn on, so there is no danger of radiation exposure to a patient."
When I asked him about the limit switches it turns out they are read by software only and the software will turn off power to the motor controllers if a limit switch is activated.
I asked why he does not wire the switches to cut power directly to be on the safe side.
His answer "It's to much bother to add the extra circuits."
We are talking less than $20 in parts and a day of his time. If the software fails after sending the controller a message to start moving the head at a certain speed then crashes there is nothing to stop the machine wreaking itself.
E.C.P.
Limit switches typically include two trip points. The first is monitored by the control system; when it is tripped the control halts execution and stops the machine. The second limit is wired directly to the servo amplifier so that, if for what ever reason, the control fails to halt the machine when the soft limit is tripped power is removed and motion is halted. Both limits are fail-safe such that if they were to become disconnected it would result in a limit exceeded condition.
3D printers are like this too. They have mechanical limit switches [0] that are read only by software. So if there is a bug in the software, nothing is stopping it from pushing the hardware limits and breaking. Same goes the other way around, if this switch is broken, same might happen.
[0] https://i.ebayimg.com/images/g/EYAAAOSwbopZguz4/s-l300.jpg
I'm much more worried about the heating element. Its temperature is usually controlled by the same cpu that also does motion control and g-code parsing. If anything locks up the CPU the heat might not be turned off in time, and (because you also want fast startup) there is enough power available to melt something. At the very least you would get nasty fumes from over-heated plastics, and maybe even teflon tape, which often is part of the print head. At worst it could start a fire.
The straightforward way of implementing this with a bidirectional motor is to wire normally-closed limit switches in series with their appropriate direction signals, such that when the switch is actuated it prevents the motor from going in that direction, but it can still move away from the switch.
It's in the article you're commenting on.