The trick with these puppies is to read the resistance without heating the thing up. We'd typically use a brief 4-probe DC measurement (4 probes so the leads don't contribute to the resistance), and inject a few or tens of nA, and read the voltage (in range of uV to mV).
Sometimes we used a lock-in amplifier, which is a pretty ingenius invention. You input a small AC signal, and then bandpass the output at the same frequency. That really cuts down on the noise at other frequencies, and lets you zero in on tiny signals from tiny currents with minimal heating.
I wondered why the thermocouple always returned to the same "room temperature" after half an hour or so, even though I and everyone else began transpiring heavily. It took me embarrassingly long to realize that the thermocouple only gives you the temperature difference between the two points of contact (tip and inside multimeter), and the multimeter took half an hour or so to warm up.
Remark: The external temperature probe is always called "hot", the reference junction always "cold", even if you measure liquid nitrogen with your sensor...
If the "cold junction" is just the banana-plugs, then it's obviously not too easy to get a reading of the temperature there. Better devices will use these rectangular thermocouple-sockets (a rectangular block with two small, dissimilar, slots) and have proper NiCr-Ni cabling inside, just up to the termination point on the PCB, right adjacent to the voltage amplifier/sensor for the reference temperature.