By adding a third electrode to replace the oxygen every time one is smashed, you maintain a perfect balance and eliminate that suction. Because the room stays full, the sensor no longer relies on the speed of the oxygen rushing in; it simply measures the steady state of the oxygen already there. Even if gunk gets on the window, the sensor won't be starved of a reading. It might take a few extra seconds for the levels to settle, but the final number will be 100% accurate because the sensor is no longer emptying its own room to get a count.
The big gain comes from a change in how you interpret the presence of electrons.
The older approach converted oxygen to electrical current, the magnitude of current flow relating to magnitude of oxygen depletion. The assumption built into that approach is that low oxygen depletion levels meant low oxygen levels, but that wasn't the only potential cause, because it ignored variation in the permiability of the membrane.
The newer approach equates current flow to oxygen concentration, as the system doesn't deplete the concentration any longer. The permiability of the membrane in this setup only contributes to a longer initial delay as the inner chamber comes to equilibrium with the surrounding concentration.
What does "adds back an oxygen molecule" mean?
Elaborate and you'll find the issue with this setup.
1: as in, one with detailed knowledge in some specialized field (as of science or literature)
Mainly I take issue with the person I replied to implying "I don't understand the solution, therefore there must be some functional issue with the solution."
How do you know when you have to do it? The sensor tells you how many oxygen molecules you consumed, as a proportion of the current flowing. So just let oxygen flow into the tank at the same rate as you're consuming it. Which you know because the device literally measures how much oxygen it is consuming.
I think the real issue is that the explanation in the tweet is from a physics perspective rather than an engineering one, which means it reads like it was implemented with impossible magic.
Mega LMAO. I can assure you this is not what's going on, at all. Also, if you release oxygen in gas form into the liquid you're going to run into a zillion other problems.
One of the golden premises of measuring things is to avoid altering what you're measuring, lol.
The point is this reaction is reversible. In one direction, you end up with fewer O2 molecules than you had before. In the other direction, you end up with more.
By keeping the state of oxygen inside the probe constant and replacing consumed molecules you now can measure almost instantly.
Edit: I think I get it now, it's a chemical reaction. By applying a voltage with some polarity to the 3rd electrode you can run the reaction in reverse. Still very hard to achieve because you have to make sure the reactions happen at the same rate with the same efficiency, which is far from trivial. This must be a very high end sensor for all this effort to make sense.
The sensors must be consumable with a certain lifetime.
Zinc can do this too. But I like silver, its oxide has decent conductivity.
One of the common arrangements on a basic two-electrode sensor is to have one gold electrode to make contact with the electrolyte, and the electrolyte provides conductivity to a sacrificial silver electrode. With electrolyte exposed to the atmosphere through an oxygen-permeable membrane.
As oxygen makes its way through the membrane, it is consumed by the silver at a steady rate and equilibrium is achieved relative to how much oxygen is in the atmosphere. This generates a steady current which is amplified to move a needle on a gauge, where there are knobs to adjust the meter until it displays the correct amount of oxygen during calibration against a known concentration. And must also be calibrated to display zero when there is no oxygen.
Eventually even if the membrane never gets fouled the oxidized silver builds up in the electrolyte chamber and response deteriorates so maintenance is needed. Remove the membrane, polish the silver, put in fresh electrolyte, new membrane, and re-calibrate.
Adding a third electrode opens up a number of further possibilities.
One of them is the option to use an additional inert gold or platinum contact or a salt bridge as electrical reference against the original gold or silver as the sensor. Plus using a more complex circuit than a plain amplifier, apply controlled responsive current to the sacrificial silver at the same time. So rather than directly amplifying the current produced by different concentrations of oxygen existing in the electrolyte (and waiting for it to equilibrate), instead with 3 (or 4) electrodes the ionic silver concentration in the electrolyte can be maintained electronically in a steady state, and as oxygen permeates, the current required to replace the consumed silver is designed to make a dfferent kind of meter move the needle the same way as above. In this way the oxygen concentration in the electrolyte varies to a much more limited extent compared to waiting for it to be depleted from a high amount to zero before the meter will bottom out.
This can be equivalent to constant-ion electrochemical titration.
Disclaimer: I always like to handle things like this like lives depended on it, because lives depended on it.
You do the best you can. If you can only make inaccurate sensors, make inaccurate sensors. If you can make accurate sensors, make accurate sensors. If they're much more expensive, make both. If your competitor has more accurate sensors, learn how they work.
(A quick google brings up this document which describes the principle. No idea if this is the company in the story: https://semeatech.com/uploads/Tech_Docs/AN%20161205.pdf )
At least that's what I assume.
Before, you measured diffusion rate of oxygen and inferred oxygen concentration from that (the concentration outside the chamber is always greater than the concentration inside). Dirty membranes etc all changed the rate of diffusion, which caused issues.
After you measure oxygen concentration directly (the concentration inside and outside the chamber are always the same).