What does the back to back ± mean? Is that the variance of accuracy from device to device? Or does the 5% reference the specific range of 400-3000?
What does the back to back ± mean? Is that the variance of accuracy from device to device? Or does the 5% reference the specific range of 400-3000?
No, it's stacked, and strictly applicable between 400-3000 ppm as characterized (despite 0-32000 ppm operating range by design).
At the lower 400 ppm end, accuracy is +/- 50 ppm +/- 20 ppm = +/- 70 ppm.
At the upper 3000 ppm end, accuracy is +/- 50 ppm +/- 150 ppm = +/- 170 ppm.
Also worth noting that the calibrated reference used to determine this published accuracy has an uncertainty of +/- 2%; ref datahseet[1] Table 7.
Caveat emptor: this published accuracy can't be trusted at face value when sampling more than 1 meas/min; ref ibid.[1] Table 4.
[1] https://www.infineon.com/dgdl/Infineon-PASCO2V15-DataSheet-v...