You not only describe how you measured your drug concentration, but you also test the limits. And those limits should go beyond anything you might encounter in actual manufacturing.
If your aim is to confirm a concentration of 5mg/mL, then you test all the way down to 0mg/mL and then 50mg and 500mg/mL. Then if you have a saline mixture, you test different concentrations of NaCl.
You’re basically trying to break the analysis to show where it doesn’t work.
The fact they did a UV vis and didn’t account for a different emulsifier is a big miss.
It has been! But if my lab experience was any guide, it's not necessarily a good light sensor. Spectrophotometry is great for getting coarse accuracy results out of. True precision... that is, and will remain, difficult.