> The Therac-25 software errors that cause radiation overexposures can be reduced down to interface errors. The first of these errors involved the entering of treatment data by the machine operator. Once an operator enters treatment information at the terminal outside of treatment room, the magnets used to filter and control radiation levels are set. There are several magnets, and the process takes about 8 seconds. If the operator makes a very, very quick change of the treatment information, within 1 second, the change is registered. Or, if the operator is rather slow about it, takes more than 8 seconds, the change is also registered. However, if the change occurs within the eight seconds it takes to set the magnets, the change is not detected and the magnets continue to be set up improperly, and thus the level of radiation is set up improperly.
> The last of the accidents occurred at the Yakima Valley Memorial Hospital. On January 17, 1987 an operator placed a patient on the turntable in the field-light position for small position verification doses. After attempting to administer the treatment dose, the machine shut down with a quick malfunction message and a treatment pause. The operator pushed the "P" button, and the machine paused again. The machine indicated that the patient had received his prescribed 7 rad of treatment. The patient, however, complained of a "burning sensation" and died three months later from complications related to the overdose (Leveson and Turner, 1993, p. 33) .
http://users.csc.calpoly.edu/~jdalbey/SWE/Papers/THERAC25.ht...