You have to be careful with isolation transformers because there is one gotcha... the ground pin normally passes through to earth. I've witnessed someone blow up a nice $5000 Agilent oscilloscope before I could stop them because they didn't understand how to use an IT and plugged the
scope into it so they could probe AC. Bad news, you're supposed to plug your test device into the IT not the scope because the BNC connectors in almost all oscilloscopes are connected straight through to the earth pin.
I heard "POW" from across the lab as soon as he clipped the probe on and knew I should have warned him. Nobody was hurt apart from our lab budget.